Details
Details:
Usage/Application AERB Approved X Ray Quality Assurance Testing Type QA Testing Measurements As Per AERB Guidelines- Complete Fitness Testing Strength AERB Approved RSO Conduct Quality Assurance Testing Service Location UP,MP,HP,Punjab,Delhi,Haryana,Bihar,Chhatisgarh,Jharkhand,Uttarakhand,Entire India Service Mode Onsite And On Line Country of Origin Made in India AERB Quality Assurance Testing for X Ray
We are AERB Authorized Agency for Performing the AERB Quality Assurance Audit process for all type of Radiation Equipment. We serve across India. We have operation base / branches in Rajasthan, Uttar Pradesh, Bihar, Telangana & Maharashtra. We have tam of AERB certified RSO (Radiation Safety Officers), who perform Quality Assurance audit cum testing of Diagnostic X ray equipment.
1. Onsite Quality Assurance Testing: AERB Approved RSO visit Facility – hospital, Diagnostic Centre & any institute using X Ray equipment. Using AERB Approved Quality Assurance Testing machine, our QA engineer conduct Quality Assurance Test of all X ray radiation generating equipment.
Quality Assurance testing of Diagnostic X-ray machines at the center as per the AERB Guideline. Following Tests to be carried out-
- Congruence Test
- Measurement of KVP ( kV testing on steps / stations )
- Linearity of mA Loading Stations ( mA testing on steps/ stations)
- Linearity of Timer
- Measurement of Exposure time
- Total Filtration
- Tube & Collimator Leakage Test ( Radiation Leakage testing to check unwanted exposure to Radiographer
- Focal Spot Test
- Measurement of Dose
- Central Beam Alignment Test
- Area Surveillance tes
2. Quality Assurance Testing Report- There are two stages in which we provide QA testing report. Online and offline.
- Offline QA testing report : Within 7 days from the date of QA testing, original QA testing report is sent to institute’s address, which is required for record to show NABL, NABH and AERB official on inspection.
- Online QA Testing reporting to AERB: Star Nuke submit QA testing report to AERB through online portal-eLORA. This is to certify that Star Nuke has carried out QA Testing of X Ray machine and X Ray machine has passed all test so comply with AERB safety standards. This Quality Assurance Report is valid for 2 years. Next QA testing is done after two years.
Additional Information:
- Production Capacity: 50 units per day
- Delivery Time: 7 Days from the date of PO
- Packaging Details: On site QA Testing followed by QA report submission to AERB online.